4 Preparing the <strong>Stabil</strong>-<strong>Ion</strong> System for Operation<br />
RS-232<br />
ELECTROMETER<br />
CONVECTRON<br />
PROCESS CONTROL<br />
IG-1 AUTO<br />
Off SP On<br />
Data Port<br />
IG1 IG2 Scale<br />
Factor<br />
Adjust<br />
Pressure<br />
Range<br />
L<br />
H L<br />
H<br />
Cathode<br />
1<br />
2 1 2<br />
IG1 IG2<br />
OFF<br />
SET<br />
ATM<br />
VAC<br />
IG-2 AUTO<br />
OFF<br />
SET<br />
ATM<br />
VAC<br />
1<br />
2<br />
3<br />
4<br />
5<br />
6<br />
3<br />
▲<br />
Set<br />
▲<br />
C<br />
H<br />
A<br />
N<br />
N<br />
E<br />
L<br />
IG1<br />
O<br />
Degas<br />
O<br />
IG2<br />
O<br />
On<br />
O<br />
Remote<br />
Figure 4-1<br />
Control Unit (with Door Open) Showing Electrometer Module Controls.<br />
a. SCALE FACTOR: Assuming the gas in the system is air or N 2 , adjust IG1 and IG2 scale<br />
factors to 1.00 with the scale factor switch set in the respective positions. Then set the<br />
scale factor switch at OFF. See Section 4.1.2 on page 4-3 for other gases.<br />
b. CATHODE: Set at 1 for primary cathode (set at 2 for alternate cathode if required).<br />
Separate switches are provided for IG1 and IG2.<br />
c. PRESSURE RANGE: Select H or L. Separate switches are provided for IG1 and IG2.<br />
The H range is preferred for applications involving pressure measurement above 1x10 -4 Torr.<br />
The emission current and cathode power are reduced and the upper pressure limit and<br />
overpressure IG shut down setpoint are extended. The H range provides increased cathode life<br />
when the IG is operated at higher pressures.<br />
The L range is preferred for applications involving pressure measurement below approximately<br />
1x10 -7 Torr. It provides the fastest response to pressures below 1x10 -7 Torr and allows<br />
measurement to the x-ray limit of the <strong>Stabil</strong>-<strong>Ion</strong> Gauge.<br />
Whether H or L is preferred between 10 -4 and 10 -7 Torr depends on whether the expected<br />
pressures will be above 10 -4 Torr or below 10 -7 Torr.<br />
The <strong>Stabil</strong>-<strong>Ion</strong> Gauge will provide the best stability and lifetime when the pressure range<br />
selector switch is set as shown in Table 4-1.<br />
4-2<br />
<strong>Series</strong> <strong>360</strong> <strong>Stabil</strong>-<strong>Ion</strong> December, 2001