4 Preparing the <strong>Stabil</strong>-<strong>Ion</strong> System for Operation<br />

RS-232<br />

ELECTROMETER<br />

CONVECTRON<br />

PROCESS CONTROL<br />

IG-1 AUTO<br />

Off SP On<br />

Data Port<br />

IG1 IG2 Scale<br />

Factor<br />

Adjust<br />

Pressure<br />

Range<br />

L<br />

H L<br />

H<br />

Cathode<br />

1<br />

2 1 2<br />

IG1 IG2<br />

OFF<br />

SET<br />

ATM<br />

VAC<br />

IG-2 AUTO<br />

OFF<br />

SET<br />

ATM<br />

VAC<br />

1<br />

2<br />

3<br />

4<br />

5<br />

6<br />

3<br />

▲<br />

Set<br />

▲<br />

C<br />

H<br />

A<br />

N<br />

N<br />

E<br />

L<br />

IG1<br />

O<br />

Degas<br />

O<br />

IG2<br />

O<br />

On<br />

O<br />

Remote<br />

Figure 4-1<br />

Control Unit (with Door Open) Showing Electrometer Module Controls.<br />

a. SCALE FACTOR: Assuming the gas in the system is air or N 2 , adjust IG1 and IG2 scale<br />

factors to 1.00 with the scale factor switch set in the respective positions. Then set the<br />

scale factor switch at OFF. See Section 4.1.2 on page 4-3 for other gases.<br />

b. CATHODE: Set at 1 for primary cathode (set at 2 for alternate cathode if required).<br />

Separate switches are provided for IG1 and IG2.<br />

c. PRESSURE RANGE: Select H or L. Separate switches are provided for IG1 and IG2.<br />

The H range is preferred for applications involving pressure measurement above 1x10 -4 Torr.<br />

The emission current and cathode power are reduced and the upper pressure limit and<br />

overpressure IG shut down setpoint are extended. The H range provides increased cathode life<br />

when the IG is operated at higher pressures.<br />

The L range is preferred for applications involving pressure measurement below approximately<br />

1x10 -7 Torr. It provides the fastest response to pressures below 1x10 -7 Torr and allows<br />

measurement to the x-ray limit of the <strong>Stabil</strong>-<strong>Ion</strong> Gauge.<br />

Whether H or L is preferred between 10 -4 and 10 -7 Torr depends on whether the expected<br />

pressures will be above 10 -4 Torr or below 10 -7 Torr.<br />

The <strong>Stabil</strong>-<strong>Ion</strong> Gauge will provide the best stability and lifetime when the pressure range<br />

selector switch is set as shown in Table 4-1.<br />

4-2<br />

<strong>Series</strong> <strong>360</strong> <strong>Stabil</strong>-<strong>Ion</strong> December, 2001