Chapter 4<br />

4 Preparing the <strong>Stabil</strong>-<strong>Ion</strong> System for<br />

Operation<br />

4.1 Preparing for Pressure Measurement with the <strong>Stabil</strong>-<strong>Ion</strong> Gauge 1<br />

1. The steps in this Section (4.1) assume:<br />

a. Your <strong>Stabil</strong>-<strong>Ion</strong> System has been properly set up and installed per the instructions in<br />

Chapters 2 and 3.<br />

b. The gas in your vacuum system is air or N 2 . For other gases you must follow the<br />

instructions in Section 4.1.3 on page 4-5 for using <strong>Stabil</strong>-<strong>Ion</strong> Gauges and in Section 4.6.1<br />

on page 4-12 for using Convectron Gauges.<br />

c. That you are reasonably familiar with the general theory of operation of hot cathode<br />

ionization gauges and thermal conductivity gauges.<br />

We recommend you consult a good text book if you are unfamiliar with vacuum technology or<br />

require more information on the general theory of operation of an ionization gauge or thermal<br />

conductivity gauge. Extremely useful information is provided in the following references.<br />

■<br />

Dushman, S., Lafferty, J. M., Scientific Foundations of <strong>Vacuum</strong> Technique, John Wiley & Sons,<br />

Inc., Second Edition, New York, 1962.<br />

■ Redhead, P. A., et al., Ultrahigh <strong>Vacuum</strong>, Chapman and Hall, London, 1968.<br />

■ O'Hanlon, J. F., A User’s Guide to <strong>Vacuum</strong> Technology, John Wiley & Sons, New York, 1980.<br />

2. Turn on the <strong>Stabil</strong>-<strong>Ion</strong> System by pressing the Control Unit power switch labeled ON. See<br />

Figure 4-1.<br />

3. Convectron Gauge equivalent N 2 pressures will be displayed whenever power is applied if the<br />

gauges and cables are installed. Make sure the IG1 AUTO and IG2 AUTO switches are in the<br />

IG AUTO ON or OFF position and have not been left in the set positions. See Section 4.6 on<br />

page 4-11 for information on Convectron Gauge pressure measurement.<br />

4. Open the front door of the Control Unit and make the following settings on the electrometer<br />

module for each <strong>Stabil</strong>-<strong>Ion</strong> Gauge channel.<br />

1. For completeness, all available capabilities are assumed present. Not all capabilities may be present in the<br />