2 Initial Setup Procedures<br />

2.1.5.1 Selecting the Byte Format<br />

Dip switches 6-8 may be used to set the baud rate. The settings are:<br />

Table 2-5<br />

Baud Rate.<br />

X<br />

X<br />

OFF<br />

ON<br />

S6 S7 S8 Baud Rate<br />

On On On 9600*<br />

On On Off 4800<br />

On Off On 2400<br />

On Off Off 1200<br />

Off On On 600<br />

Off On Off 300<br />

Off Off On 150<br />

Off Off Off 75<br />

* factory setting<br />

Figure 2-14<br />

Partial Top View of Process Control Module.<br />

2.1.5.1.1 Character Framing<br />

Switches 3-5 control the number of characters, polarity, and number of stop bits:<br />

Table 2-6 Character Framing.<br />

S3 S4 S5<br />

Character<br />

Bits Parity Stop Bits<br />

X<br />

X<br />

OFF<br />

ON<br />

On On On 8 None 2<br />

On On Off 8 Even 2<br />

On Off On 8 Odd 1<br />

On* Off* Off* 8* None* 1*<br />

Off On On 7 Even 1<br />

Off On Off 7 Odd 1<br />

Off Off On 7 Even 2<br />

Off Off Off 7 Odd 2<br />

* factory setting<br />