Series 360, Granville-Phillips, Stabil-Ion, Ionization, Vacuum ...
Series 360, Granville-Phillips, Stabil-Ion, Ionization, Vacuum ...
Series 360, Granville-Phillips, Stabil-Ion, Ionization, Vacuum ...
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
2 Initial Setup Procedures<br />
2.1.5.1 Selecting the Byte Format<br />
Dip switches 6-8 may be used to set the baud rate. The settings are:<br />
Table 2-5<br />
Baud Rate.<br />
X<br />
X<br />
OFF<br />
ON<br />
S6 S7 S8 Baud Rate<br />
On On On 9600*<br />
On On Off 4800<br />
On Off On 2400<br />
On Off Off 1200<br />
Off On On 600<br />
Off On Off 300<br />
Off Off On 150<br />
Off Off Off 75<br />
* factory setting<br />
Figure 2-14<br />
Partial Top View of Process Control Module.<br />
2.1.5.1.1 Character Framing<br />
Switches 3-5 control the number of characters, polarity, and number of stop bits:<br />
Table 2-6 Character Framing.<br />
S3 S4 S5<br />
Character<br />
Bits Parity Stop Bits<br />
X<br />
X<br />
OFF<br />
ON<br />
On On On 8 None 2<br />
On On Off 8 Even 2<br />
On Off On 8 Odd 1<br />
On* Off* Off* 8* None* 1*<br />
Off On On 7 Even 1<br />
Off On Off 7 Odd 1<br />
Off Off On 7 Even 2<br />
Off Off Off 7 Odd 2<br />
* factory setting<br />
Figure 2-15<br />
Partial Top View of RS-232 Module.<br />
2-10<br />
<strong>Series</strong> <strong>360</strong> <strong>Stabil</strong>-<strong>Ion</strong> December, 2001