2 Initial Setup Procedures<br />

2.1.4 Process Control Setup<br />

It is the installer's responsibility to ensure that the automatic signals<br />

provided by the product are always used in a safe manner. Carefully check<br />

the system programming before switching to automatic operation.<br />

Where an equipment malfunction could cause a hazardous situation, always<br />

provide for fail-safe operation. As an example, in an automatic backfill<br />

operation where a malfunction might cause high internal pressures, provide<br />

an appropriate pressure relief device.<br />

1. If you do not have process control capability skip to Section 2.1.5 on page 2-9.<br />

2. Use the catalog number on the front of the Process Control module together with Figures. 2-9,<br />

2-10. and 2-11 to identify the process control capability installed in your unit. For Cat. No.<br />

<strong>360</strong>109, only the channels you have ordered will have switches present on the front of the<br />

module., e.g., 1, 2, 3, or 5.<br />

3. Even if the control logic is simple and obvious, we recommend that you develop a logic<br />

diagram of the process control function.<br />

4. Prepare a specification table which lists the proposed pressure setting, system measurement<br />

point, and relay status for each Process Control channel.<br />

5. Draw a circuit schematic which specifies exactly how each piece of system hardware will be<br />

connected to the Process Control relays.<br />

Do not exceed the relay ratings:<br />

Table 2-1<br />

Relay Ratings.<br />

Relay Configuration<br />

Relay Contact Rating<br />

Relay Contact Type<br />

SPDT (single pole, double throw)<br />

5 A, 60 Vdc resistive load, 30 V rms (42.4 Vpk)<br />

5 A (gold plated for low level switching)<br />

NOTE: If the relay contacts are used to switch high currents, the gold plating may be consumed.<br />

This may make the contacts unsuitable for low level signal switching in the future.<br />

6. Attach a copy of the Process Control circuit diagram to this manual for future reference and<br />

troubleshooting.<br />

7. The required process control connections may be made later. See Section 3.6 on page 3-8.<br />

8. If application assistance is desired, contact a Helix Technology Application Engineer at the<br />

number listed on the title page of this manual.<br />