<strong>On</strong>-Board W <strong>Module</strong> <strong>Programming</strong> <strong>and</strong> <strong>Operation</strong><br />

CTI-CRYOGENICS<br />

HELIX TECHNOLOGY CORPORATION<br />

Appendix E - <strong>On</strong>-Board Pump<br />

Fault Messages<br />

Introduction<br />

Appendix E describes the <strong>On</strong>-Board pump fault messages which are<br />

requested by host computer comm<strong>and</strong>s. The fault message indicates that a<br />

problem exists within the pump.<br />

Table E-1: Host Comm<strong>and</strong>s <strong>and</strong> Fault Messages<br />

Host Comm<strong>and</strong> Action Description<br />

S3<br />

S3<br />

v<br />

v<br />

Logical AND the<br />

response with 01H<br />

Logical AND the<br />

response with02H<br />

Logical AND the<br />

response with 04H<br />

Logical AND the<br />

response with 01H<br />

A result of 1 means one phase of pump power is<br />

missing.<br />

A result of 1 means both phases of pump power<br />

are missing.<br />

When TRUE, a heater failure has been detected<br />

<strong>and</strong> the warm-up temperature is reduced to 285K.<br />

When TRUE, there is a contention for the<br />

roughing valve.<br />