Appendix C - RS-232 REGEN Responses<br />
CTI-CRYOGENICS<br />
HELIX TECHNOLOGY CORPORATION<br />
Table C-1: RS-232 REGEN Responses <strong>and</strong> Descriptions<br />
REGEN Response<br />
A<br />
D<br />
Cryopump OFF<br />
Description<br />
WARM-UP - During warm-up you may also want to display<br />
the actual 1st stage temperature <strong>and</strong> the temperature it is attempting to<br />
reach. The reduced temperature of 285K represents a failure of the<br />
pump to reach 310K in 60 minutes.<br />
Example: WARM-UP 272K/285K<br />
The 285 is fixed <strong>and</strong> the 272 is the response from the<br />
comm<strong>and</strong> K.<br />
E<br />
WARM-UP - During warm-up you may also want to display<br />
the actual 1st stage temperature <strong>and</strong> the temperature it is attempting to<br />
reach.<br />
Example: WARM-UP 234K/310K<br />
The 310K is fixed <strong>and</strong> the 234 is the response from the<br />
comm<strong>and</strong> K.<br />
H<br />
EXTENDED PURGE<br />
An H response means that the pump has reached its warm-up temperature<br />
<strong>and</strong> the purge is being extended for some amount of time. The<br />
time is determined when the regeneration parameters are<br />
programmed. The time remaining in minutes is comm<strong>and</strong> k.<br />
Example: EXTENDED PURGE<br />
12 MIN<br />
C-2 P/N 8040365