Appendix C - RS-232 REGEN Responses<br />

CTI-CRYOGENICS<br />

HELIX TECHNOLOGY CORPORATION<br />

Table C-1: RS-232 REGEN Responses <strong>and</strong> Descriptions<br />

REGEN Response<br />

A<br />

D<br />

Cryopump OFF<br />

Description<br />

WARM-UP - During warm-up you may also want to display<br />

the actual 1st stage temperature <strong>and</strong> the temperature it is attempting to<br />

reach. The reduced temperature of 285K represents a failure of the<br />

pump to reach 310K in 60 minutes.<br />

Example: WARM-UP 272K/285K<br />

The 285 is fixed <strong>and</strong> the 272 is the response from the<br />

comm<strong>and</strong> K.<br />

E<br />

WARM-UP - During warm-up you may also want to display<br />

the actual 1st stage temperature <strong>and</strong> the temperature it is attempting to<br />

reach.<br />

Example: WARM-UP 234K/310K<br />

The 310K is fixed <strong>and</strong> the 234 is the response from the<br />

comm<strong>and</strong> K.<br />

H<br />

EXTENDED PURGE<br />

An H response means that the pump has reached its warm-up temperature<br />

<strong>and</strong> the purge is being extended for some amount of time. The<br />

time is determined when the regeneration parameters are<br />

programmed. The time remaining in minutes is comm<strong>and</strong> k.<br />

Example: EXTENDED PURGE<br />

12 MIN<br />

C-2 P/N 8040365