<strong>On</strong>-Board W <strong>Module</strong> <strong>Programming</strong> <strong>and</strong> <strong>Operation</strong><br />

CTI-CRYOGENICS<br />

HELIX TECHNOLOGY CORPORATION<br />

Table B-1: <strong>On</strong>-Board <strong>Module</strong> RS-232 Comm<strong>and</strong>s (Continued)<br />

Comm<strong>and</strong> Parameter Description<br />

T<br />

[1-2] {0-8,ABCF]<br />

[4 digit number]<br />

{| {[?] [0,1,2,8]}<br />

This comm<strong>and</strong> programs the setpoint relays. The first parameter<br />

tells the relay number (1 or 2). The second parameter is one<br />

of two forms. The first, used for setting the item to be programmed,<br />

consists of a selector (0,1,4-8). This form is followed<br />

by a third parameter, a 4 four digit number, if the selector<br />

is 0 - 8. The A - F selectors do not require a following numeric<br />

parameter, as they are logical function settings.<br />

The second form is used to query the settings <strong>and</strong> consists of<br />

the "?" comm<strong>and</strong> element followed by a selector (0-2, 8) to<br />

indicate what data is being queried.<br />

Selector meaning for first form:<br />

0 = Set first stage LL (deg Kelvin)<br />

1 = Set first stage UL<br />

4 = Set Pump TC LL (microns)<br />

5 = Set Pump TC UL<br />

6 = Set AUX TC LL (microns)<br />

7 = Set AUX TC UL<br />

8 = Set time relay (0-9999 seconds)<br />

A = Set unconditional ON<br />

B = Set unconditional OFF<br />

C = Set as REGEN status tracking relay<br />

D = Set as rough valve tracking relay<br />

F = Set as pump motor ON/OFF status tracking relay<br />

Selector meaning for second form:<br />

0:=Query previously set LL (meaning depends upon previous use)<br />

1:=Query previously set UL (meaning depends upon previous use)<br />

2:=Query last sent first form selector (0,1,4-8) used to<br />

establish particular relay function<br />

8:=Query relay time delay (0-9999 seconds)<br />

Example: "$T 1 8 30" /*Set relay time delay to 30 seconds*/<br />

VA [?] Read pump serial number.<br />

Y ? Read elapsed time. The units are in hours <strong>and</strong> expressed as an<br />

integer. The range is up to 65,000 hours.<br />

Z [?] Read the number of REGEN cycles that have been completed.<br />