On-Board® W Module Programming and Operation ... - PTB Sales
On-Board® W Module Programming and Operation ... - PTB Sales
On-Board® W Module Programming and Operation ... - PTB Sales
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
Appendix B - <strong>On</strong>-Board <strong>Module</strong> RS-232 Interface Protocol Format<br />
CTI-CRYOGENICS<br />
HELIX TECHNOLOGY CORPORATION<br />
Table B-1: <strong>On</strong>-Board <strong>Module</strong> RS-232 Comm<strong>and</strong>s (Continued)<br />
Comm<strong>and</strong> Parameter Description<br />
S 1 - 3 Send back various status indicators.<br />
Status 1<br />
01H = Pump ON<br />
02H = Rough valve ON<br />
04H = Purge valve ON<br />
08H = TC1 ON<br />
10H = TC2 ON<br />
20H = A result of 0 means that there was a power failure. The<br />
bit is reset to 1 after this query.<br />
Status 2<br />
01H = Setpoint 1 ON<br />
08H = First stage temperature control ON<br />
Status 3<br />
01H = Pump phase 1 check<br />
02H = Pump phase 2 check<br />
B-8 P/N 8040365