Appendix B - <strong>On</strong>-Board <strong>Module</strong> RS-232 Interface Protocol Format<br />

CTI-CRYOGENICS<br />

HELIX TECHNOLOGY CORPORATION<br />

Table B-1: <strong>On</strong>-Board <strong>Module</strong> RS-232 Comm<strong>and</strong>s (Continued)<br />

Comm<strong>and</strong> Parameter Description<br />

S 1 - 3 Send back various status indicators.<br />

Status 1<br />

01H = Pump ON<br />

02H = Rough valve ON<br />

04H = Purge valve ON<br />

08H = TC1 ON<br />

10H = TC2 ON<br />

20H = A result of 0 means that there was a power failure. The<br />

bit is reset to 1 after this query.<br />

Status 2<br />

01H = Setpoint 1 ON<br />

08H = First stage temperature control ON<br />

Status 3<br />

01H = Pump phase 1 check<br />

02H = Pump phase 2 check<br />