<strong>Programming</strong> <strong>and</strong> <strong>Operation</strong><br />
CTI-CRYOGENICS<br />
HELIX TECHNOLOGY CORPORATION<br />
6. Press the MONITOR key. The <strong>On</strong>-Board microprocessor will<br />
monitor system operation.<br />
First Stage Temperature Control<br />
NOTE: The First Stage Temperature Control parameter cannot be<br />
accessed unless a password has been entered.<br />
1. Press the SERVICE key on the keypad/display. The serial<br />
number <strong>and</strong> software version information are displayed.<br />
2. Press the Next key until the Password? prompt is displayed.<br />
3. Enter the password <strong>and</strong> press Enter.<br />
4. Press the Next key until 1st Stg Temp is displayed.<br />
5. Press the Enter key. Press the numeral keys for the desired<br />
temperature setpoint. The range of values is 107K - 320K.<br />
6. Press the Enter button.<br />
7. Press the MONITOR key. The <strong>On</strong>-Board microprocessor will<br />
monitor system operation.<br />
3-38 P/N 8040365