Appendix B - <strong>On</strong>-Board <strong>Module</strong> RS-232 Interface Protocol Format<br />
CTI-CRYOGENICS<br />
HELIX TECHNOLOGY CORPORATION<br />
Table B-1: <strong>On</strong>-Board <strong>Module</strong> RS-232 Comm<strong>and</strong>s (Continued)<br />
Comm<strong>and</strong> Parameter Description<br />
@ None Returns an identifier string indicating module type <strong>and</strong> software<br />
revision level.<br />
a None Read back how long it has been since the last REGEN cycle.<br />
The units are in hours.<br />
e None Read error code result of REGEN operation. Returns one of<br />
eight characters as defined in Appendix C, should only be used<br />
when REGEN response V is returned.<br />
g None Requests the auto calibration of the pump TC gauge. Returns a<br />
"G" or "H" error code if conditions are not valid to perform the<br />
auto zero calibrate, such as TC not turned ON or the pressure is<br />
too high.<br />
h None Requests the auto calibration of the AUX TC gauge. Returns a<br />
"G" or "H" error code if conditions are not valid to perform the<br />
auto zero calibrate, such as TC not turned ON or the pressure is<br />
too high.<br />
i [0,1,2,?] Set/Clear/Query the power fail recovery flag.<br />
0 - Power Fail Recovery turned OFF<br />
1 - Power Fail Recovery turned ON<br />
2 - Power Fail Recovery if below power failure temperature<br />
limit<br />
? - Provides Power Fail Recovery status<br />
j [0..59994, ?] Pump delay to start of regeneration cycle (minutes).<br />
k None Returns an internal interval timer value in minutes. When in<br />
the appropriate phase, this value represents the time remaining<br />
in the phase (delay start, cryo restart, purge time). Times from<br />
1 to 60 seconds return as 1 minute. Times from 61 seconds to<br />
120 seconds return as 2 minutes. Zero returns as 0 minutes.<br />
l None Returns the number of failed purge cycles as an integer value.<br />
This value starts at "0" <strong>and</strong> increments each time a purge phase<br />
fails.<br />
m None Returns the number of failed <strong>and</strong> extended ROR test cycles as<br />
an integer value. This value starts at "0" <strong>and</strong> increments each<br />
time a ROR test phase fails.<br />
B-10 P/N 8040365