From page 3-22<br />

ROUGH 125µ/50µ<br />

Displays the current pressure (125µ) <strong>and</strong> the base pressure to be<br />

achieved (50µ). The waterpump is roughed out until the base<br />

pressure is reached.<br />

REPURGE #1 2M<br />

If the pressure does not reduce at a reasonable rate during the rough<br />

out phase, the <strong>On</strong>-Board W <strong>Module</strong> will automatically repurge the<br />

waterpump for the programmed REPURGE time. The display<br />

indicates that this is the first purge cycle (#1) <strong>and</strong> that there are two<br />

minutes (2M) remaining in this cycle.<br />

The waterpump is roughed out again until the base pressure is reached.<br />

The following events occur:<br />

- Waterpump TC reaches programmed base pressure of 50µ<br />

- Rough valve closes <strong>and</strong> the <strong>On</strong>-Board W <strong>Module</strong> waits 15 seconds<br />

to stabilize<br />

- <strong>On</strong>-Board W <strong>Module</strong> measures TC pressure <strong>and</strong> takes second reading<br />

in 30 seconds<br />

- <strong>On</strong>-Board W <strong>Module</strong> calculates ROR pressure increase in µ/minute<br />

- If ROR is less than programmed value, the ROR test passes<br />

- If ROR is more than programmed value, the ROR test is repeated<br />

ROR 3 124µ/10µ<br />

The display indicates the number of ROR test cycles (3), current ROR<br />

value (124µ/µ), <strong>and</strong> the programmed ROR value (10µ/µ).<br />

If ROR test is repeated:<br />

- Rough valve opens<br />

- Waterpump is roughed out again to base pressure<br />

- ROR test is repeated<br />

Continued on page 3-24<br />

Figure 3-23: <strong>On</strong>-Board Appendage Waterpump<br />

REGEN Cycle Program Example (continued)<br />

P/N 8040365 3-23