Table B-1: <strong>On</strong>-Board <strong>Module</strong> RS-232 Comm<strong>and</strong>s (Continued)<br />
Comm<strong>and</strong> Parameter Description<br />
n None Returns the last measured ROR value computed during a ROR<br />
test cycle as an integer value. This value matches that presented<br />
on the keypad/display.<br />
r<br />
P | X, form is<br />
rP or rX<br />
Polls an event count value which specifies the completion of<br />
either the pump (P) or AUX (X) TC auto zero sequence. This<br />
value is used by a remote computer to tell when an initiated<br />
auto zero is complete.<br />
Returns a modulo 256 (0-255) numeric value which is incremented<br />
each time the auto zero operation completes. The<br />
remote unit can detect the completion of the event by a change<br />
in value.<br />
s None Polls an event count modulo 256 (0-255) which is incremented<br />
each time a REGEN cycle is performed. Used to indicate<br />
whether the REGEN idle message should be REGEN COM-<br />
PLETE or REGEN OFF.<br />
t ? | =, form is t?<br />
or t=<br />
Comm<strong>and</strong> to either inquire (t?) or acknowledge <strong>and</strong> clear (t=)<br />
the power recovery flag. A response of 1 is COOLDOWN, 2 is<br />
REGENERATION, 3 is IN PROCESS, 4 is RECOVERED, 5 is<br />
CHECK PUMP TEMP. Clear the flag with the t= comm<strong>and</strong><br />
after inquiring on the flag value.<br />
u None Inquires pump for information in byte fail_cause of pump,<br />
which contains 0 if AC to pump OK, else bits 4,5 set for fail of<br />
each phase. Codes are@, P,p. Used to display of power phase<br />
information of pump. OFF, ON, ERR1, <strong>and</strong> ERR2.<br />
v None Inquires pump for several flag conditions combined as bits in<br />
one response. The bits are biased by "0" for ASCII printable<br />
results. The bit weights are:<br />
1. Contending for rough valve is TRUE.<br />
2. TRUE when purge gas fail detected in REGEN.<br />
4. TRUE when heater failure detected in REGEN.<br />
z [0,1, ?] Set/Clear/Query the user keypad access lockout state. When<br />
True, the user is limited to the use of the MONITOR <strong>and</strong><br />
SERVICE functions on the keypad/display.<br />
P/N 8040365 B-11