10.04.2014 Views

On-Board® W Module Programming and Operation ... - PTB Sales

On-Board® W Module Programming and Operation ... - PTB Sales

On-Board® W Module Programming and Operation ... - PTB Sales

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

<strong>On</strong>-Board W <strong>Module</strong> <strong>Programming</strong> <strong>and</strong> <strong>Operation</strong><br />

CTI-CRYOGENICS<br />

HELIX TECHNOLOGY CORPORATION<br />

Table C-1: RS-232 REGEN Responses <strong>and</strong> Descriptions (Continued)<br />

REGEN Response<br />

V<br />

REGENERATION ABORTED<br />

Description<br />

Comm<strong>and</strong> e displays the reasons for regeneration abort. The<br />

messages are as follows:<br />

@ - No error<br />

A,B - WARM-UP TIMEOUT - Did not reach warm-up temperature<br />

within 150 minutes.<br />

C - COOLDOWN TIMEOUT - Did not cool down within 4 hours.<br />

E - ROR CYCLE LIMIT WAS REACHED - The pump did<br />

not pass the rate-of-rise test within the programmed cycle limit.<br />

F - MANUAL ABORT - Someone manually gave the<br />

comm<strong>and</strong> to stop regeneration.<br />

G - ROUGH VALVE TIMEOUT - Rough valve was opened<br />

for more than 1 hour without closing.<br />

W,p<br />

DELAY RESTART<br />

This countdown timer delays starting the pump for cooldown mode<br />

after it has been warmed up, roughed out, <strong>and</strong> passed the ROR test.<br />

This is a user programmed value. If the pump has no valves, the delay<br />

timer begins when the pump reaches 310K.<br />

To get the timer value in minutes, use the comm<strong>and</strong> k. Display the<br />

time in hours.<br />

X<br />

POWER FAILURE<br />

There was a power failure during the regeneration cycle <strong>and</strong> the pump<br />

will make a decision to cooldown or regenerate. This display should<br />

last for no longer than one second.<br />

Y<br />

POWER FAILURE COOLDOWN<br />

There was a power failure during regeneration but the pump is continuing<br />

to cool down. Refer to Appendix D - Power Failure<br />

Messages.<br />