08.11.2014 Views

Inogen One G3 Technical Manual

Inogen One G3 Technical Manual

Inogen One G3 Technical Manual

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

2. Seat the feet of the front housing into the corresponding grooves on the chassis (Figure 4)<br />

then push down and toward the device to secure the housing in the chassis.<br />

Figure 4: Housing feet and corresponding chassis grooves<br />

3. Replace the two screws that connect the front housing to the chassis (Figure 2). Tighten<br />

screws to 6 in-lbs.<br />

4. Return the concentrator to the upright position<br />

5. Slide the battery back into the chassis. Ensure the battery clip is fully seated.<br />

96-03997-00-01 revA Service <strong>Manual</strong>, <strong>Inogen</strong> <strong>One</strong> <strong>G3</strong><br />

- 108 -

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!