5. Plug in the compressor cable (Figure 8). Double check that the cable housing is locked<br />

into the receptacle.<br />

6. Place the black exhaust tube back through the compressor cable loop (Figure 8).<br />

7. With all cables and tubing reattached, make sure all tubes are properly oriented in their<br />

respected positions on the chassis before reinstalling the tube bracket. Reinstall the two<br />

screws (Figure 3). Tighten screws to 6 in-lbs.<br />

Instructions for Replacing the Housing:<br />

1. Seat the feet of the front housing into the corresponding grooves on the chassis (Figure<br />

11) then push down and toward the device to secure the housing in the chassis.<br />

Figure 11: Housing feet and corresponding chassis grooves<br />

2. Replace the two screws that connect the front housing to the chassis (Figure 2). Tighten<br />

screws to 6 in-lbs.<br />

3. Return the concentrator to the upright position<br />

4. Slide the battery back into the chassis. Ensure the battery clip is fully seated<br />

96-03997-00-01 revA Service <strong>Manual</strong>, <strong>Inogen</strong> <strong>One</strong> <strong>G3</strong><br />