08.11.2014 Views

Inogen One G3 Technical Manual

Inogen One G3 Technical Manual

Inogen One G3 Technical Manual

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

SP-507 <strong>Inogen</strong> <strong>One</strong> <strong>G3</strong> Filter Box Lid Replacement<br />

Parts Required:<br />

1. SP-507, Housing, Filter Box, Lid<br />

Tools Required:<br />

1. T10 Torx Driver<br />

2. Small Flathead Screwdriver<br />

Notes:<br />

1. When removing screws from the device carefully set in a bin or secure area as these<br />

screws will be used to secure the replacement parts.<br />

2. Many of the screws used on the <strong>Inogen</strong> <strong>One</strong> <strong>G3</strong> have pre-applied loctite to the threads.<br />

These loctite screws can be re-used up to four times before replacement screws are<br />

required (If exceeded four uses please contact <strong>Inogen</strong> for a screw replacement kit).<br />

Instructions for Removing the Filter Box Lid:<br />

1. While facing the front housing, remove the battery from the concentrator by pushing<br />

down on the blue battery release clip and sliding the device away (Figure 1).<br />

Figure 1: Direction to slide device to remove battery<br />

2. Place the device on its back and remove the four screws connecting the filter box lid to<br />

the device (Figure 2).<br />

96-03997-00-01 revA Service <strong>Manual</strong>, <strong>Inogen</strong> <strong>One</strong> <strong>G3</strong><br />

- 34 -

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!