Inogen One G3 Technical Manual
Inogen One G3 Technical Manual
Inogen One G3 Technical Manual
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
SP-507 <strong>Inogen</strong> <strong>One</strong> <strong>G3</strong> Filter Box Lid Replacement<br />
Parts Required:<br />
1. SP-507, Housing, Filter Box, Lid<br />
Tools Required:<br />
1. T10 Torx Driver<br />
2. Small Flathead Screwdriver<br />
Notes:<br />
1. When removing screws from the device carefully set in a bin or secure area as these<br />
screws will be used to secure the replacement parts.<br />
2. Many of the screws used on the <strong>Inogen</strong> <strong>One</strong> <strong>G3</strong> have pre-applied loctite to the threads.<br />
These loctite screws can be re-used up to four times before replacement screws are<br />
required (If exceeded four uses please contact <strong>Inogen</strong> for a screw replacement kit).<br />
Instructions for Removing the Filter Box Lid:<br />
1. While facing the front housing, remove the battery from the concentrator by pushing<br />
down on the blue battery release clip and sliding the device away (Figure 1).<br />
Figure 1: Direction to slide device to remove battery<br />
2. Place the device on its back and remove the four screws connecting the filter box lid to<br />
the device (Figure 2).<br />
96-03997-00-01 revA Service <strong>Manual</strong>, <strong>Inogen</strong> <strong>One</strong> <strong>G3</strong><br />
- 34 -