Figure 16: Housing feet and corresponding chassis grooves<br />
7. Place the device on its front side and repeat step 2 to secure the back housing into the<br />
chassis.<br />
8. Replace the two screws that connect the back housing to the chassis and the two screws<br />
that connect the front housing to the chassis (Figures 2 and 4). Tighten screws to 6 in-lbs.<br />
9. Return the concentrator to the upright position<br />
10. Slide the battery back into the chassis. Ensure the battery clip is fully seated.<br />
96-03997-00-01 revA Service <strong>Manual</strong>, <strong>Inogen</strong> <strong>One</strong> <strong>G3</strong><br />