Figure 6: Location of front screws<br />

2. Gently pull upward on the user interface panel to lift it about a centimeter off the device,<br />

being careful not to pull on the cables that are connected to the user interface panel. With<br />

the UIP cable connector exposed on the motherboard, gently unplug the UIP cable from<br />

the motherboard (Figure 7).<br />

Figure 7: Unplug UIP cable from motherboard<br />

3. Gently unplug the cannula tube and the LCD cable from the user interface panel (Figure<br />

8). Remove the user interface panel.<br />

96-03997-00-01 revA Service <strong>Manual</strong>, <strong>Inogen</strong> <strong>One</strong> <strong>G3</strong><br />