SP-501 and SP-502 <strong>Inogen</strong> <strong>One</strong> <strong>G3</strong> Front and Rear Housing<br />

Replacement<br />

Parts Required:<br />

1. SP-501, <strong>G3</strong> Housing, Front<br />

2. SP-502, <strong>G3</strong> Housing, Rear<br />

Tools Required:<br />

1. T10 Torx Driver<br />

Notes:<br />

1. When removing screws from the device, carefully set in a bin or secure area as these<br />

screws will used to secure the replacement parts.<br />

2. Many of the screws used on the <strong>Inogen</strong> <strong>One</strong> <strong>G3</strong> have pre-applied loctite on the threads.<br />

These loctite screws can be re-used up to four times before replacement screws are<br />

required (If exceeded four uses please contact <strong>Inogen</strong> for a screw replacement kit).<br />

Instructions for Removing the Housing:<br />

1. While facing the front housing, remove the battery from the concentrator by pushing<br />

down on the blue battery release clip and sliding the device away (Figure 1).<br />

Figure 1: Direction to slide device to remove battery<br />

96-03997-00-01 revA Service <strong>Manual</strong>, <strong>Inogen</strong> <strong>One</strong> <strong>G3</strong><br />

- 3 -