O2 Delivery Error No Error Message O2 Delivery Error The system detected a breath but did not<br />
detect a bolus delivery<br />
Check for kinked tubing to the breath<br />
detect sensor. Replace the breath detect<br />
sensor or the Product manifold to<br />
resolve the no bolus condition.<br />
3 Abnormality Normal Behavior Abnormal Behavior Fault Troubleshooting & Repair<br />
Noise<br />
Concentrator should be<br />
free of rattles and<br />
whistles<br />
Concentrator makes loud<br />
noise when running<br />
Compressor tube not seated correctly<br />
Push the tubing onto the compressor<br />
output so that it is all the way flush<br />
against the end of the barb. Rotate the<br />
tubing so that the compressor does not<br />
hit the rear wall of the chassis<br />
Check that the compressor mounts are<br />
not torn<br />
Concentrator makes a<br />
whistling sound during part<br />
of the operating cycle<br />
Product manifold check valve may have<br />
debris in it or may be misassembled<br />
Replace the product manifold<br />
96-03054-00 Service <strong>Manual</strong>, <strong>Inogen</strong> <strong>One</strong> <strong>G3</strong><br />
- 153 -