08.11.2014 Views

Inogen One G3 Technical Manual

Inogen One G3 Technical Manual

Inogen One G3 Technical Manual

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

O2 Delivery Error No Error Message O2 Delivery Error The system detected a breath but did not<br />

detect a bolus delivery<br />

Check for kinked tubing to the breath<br />

detect sensor. Replace the breath detect<br />

sensor or the Product manifold to<br />

resolve the no bolus condition.<br />

3 Abnormality Normal Behavior Abnormal Behavior Fault Troubleshooting & Repair<br />

Noise<br />

Concentrator should be<br />

free of rattles and<br />

whistles<br />

Concentrator makes loud<br />

noise when running<br />

Compressor tube not seated correctly<br />

Push the tubing onto the compressor<br />

output so that it is all the way flush<br />

against the end of the barb. Rotate the<br />

tubing so that the compressor does not<br />

hit the rear wall of the chassis<br />

Check that the compressor mounts are<br />

not torn<br />

Concentrator makes a<br />

whistling sound during part<br />

of the operating cycle<br />

Product manifold check valve may have<br />

debris in it or may be misassembled<br />

Replace the product manifold<br />

96-03054-00 Service <strong>Manual</strong>, <strong>Inogen</strong> <strong>One</strong> <strong>G3</strong><br />

- 153 -

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!