Product manifold tubing<br />

Figure 10: Accumulator tubing for removal<br />

4. Remove the accumulator from the device.<br />

Instructions for Replacing the Accumulator:<br />

1. Before installing the new accumulator, string a new zip tie through the loop in the chassis<br />

as shown in figure 11.<br />

Figure 11: Zip tie properly oriented in chassis<br />

2. Plug the tubing on the new accumulator into its quick connect on the product manifold<br />

(Figure 10). Place the new accumulator in its designated space and tighten the zip tie<br />

around the accumulator so that the end of the zip tie is nearest the oxygen sensor tube<br />

away from the breath detect sensor. Cut off the excess zip tie (Figure 12).<br />

96-03997-00-01 revA Service <strong>Manual</strong>, <strong>Inogen</strong> <strong>One</strong> <strong>G3</strong><br />

- 80 -