Product manifold tubing<br />
Figure 10: Accumulator tubing for removal<br />
4. Remove the accumulator from the device.<br />
Instructions for Replacing the Accumulator:<br />
1. Before installing the new accumulator, string a new zip tie through the loop in the chassis<br />
as shown in figure 11.<br />
Figure 11: Zip tie properly oriented in chassis<br />
2. Plug the tubing on the new accumulator into its quick connect on the product manifold<br />
(Figure 10). Place the new accumulator in its designated space and tighten the zip tie<br />
around the accumulator so that the end of the zip tie is nearest the oxygen sensor tube<br />
away from the breath detect sensor. Cut off the excess zip tie (Figure 12).<br />
96-03997-00-01 revA Service <strong>Manual</strong>, <strong>Inogen</strong> <strong>One</strong> <strong>G3</strong><br />