Figure 2: Location of screws<br />

3. Slide the front housing down and away from the device then lift up to remove.<br />

Instructions for Removing the Column Receptacle:<br />

1. With the device still on its back side, unscrew the two tube bracket screws (Figure 3) and<br />

remove the tube bracket and set aside.<br />

2. Disconnect the manifold tube corresponding to the receptacle being removed. Do this by<br />

pressing down on the retaining ring while pulling gently on the tubing.<br />

Manifold tubing<br />

Figure 3: Tube bracket screws<br />

3. In order to remove the receptacle, the corresponding column must be removed. Locate<br />

the column latch on the underside of the device (Figure 4) and push downward on it<br />

while simultaneously pulling the column out of the chassis. Remove the column and set<br />

aside.<br />

96-03997-00-01 revA Service <strong>Manual</strong>, <strong>Inogen</strong> <strong>One</strong> <strong>G3</strong><br />