O2 sensor tubing<br />

Accumulator tubing<br />

Figure 12: Disconnect all tubes that connect to the motherboard<br />

5. With all cables and tubing detached, remove the motherboard from the device and set<br />

face up.<br />

6. Locate the breath detect on the motherboard and unplug its tubing from the oxygen<br />

sensor by pulling the tube off of the barb (Figure 13).<br />

Breath Detect tubing<br />

Figure 13: Disconnect tubing from oxygen sensor<br />

7. Carefully pull the breath detect off of the motherboard by lifting the white board straight<br />

up.<br />

Instructions for Replacing the Breath Detect:<br />

1. Install the new breath detect by aligning the 6 pins on the bottom of the breath detect with<br />

the corresponding holes on the motherboard (Figure 14). Make sure the pins are fully<br />

seated in the holes and the breath detect is lying flat on the motherboard (Figure 15).<br />

96-03997-00-01 revA Service <strong>Manual</strong>, <strong>Inogen</strong> <strong>One</strong> <strong>G3</strong><br />

- 89 -