2. To remove the back housing, place the device on its front side and remove the two screws<br />

that secure the back housing to the chassis (Figure 2).<br />

Figure 2: Location of screws<br />

3. With the concentrator still on its front side, slide the back housing down and away from<br />

the device then lift up to remove (Figure 3).<br />

Figure 3: Direction to slide housing<br />

4. To remove the front housing, place the device on its back side and remove the two screws<br />

that secure the front housing to the chassis (Figure 4).<br />

96-03997-00-01 revA Service <strong>Manual</strong>, <strong>Inogen</strong> <strong>One</strong> <strong>G3</strong><br />