7. Place the device on its front side and repeat step 2 to secure the back housing into the<br />

chassis.<br />

8. Replace the two screws that connect the back housing to the chassis and the two screws<br />

that connect the front housing to the chassis (Figures 2 and 4). Tighten screws to 6 in-lbs.<br />

9. Return the concentrator to the upright position<br />

10. Slide the battery back into the chassis. Ensure the battery clip is fully seated.<br />