Cannula tube<br />

LCD cable<br />

Figure 8: Remove cannula tube and power cable<br />

Instructions for Removing the Inlet T Tube:<br />

1. Remove the upper feed tube from both feed waste manifolds using a small flathead screw<br />

driver, if necessary (Figure 9). Be careful not to damage the sealing surface of the<br />

manifold barb when removing the tube.<br />

Figure 9: Inlet T Tube and feed waste manifold<br />

.<br />

2. Detach the Inlet T tube from the inlet union barb (Figure 10) and remove the Inlet T tube<br />

from the device.<br />