7. Assure both lower grommets are in place against the receptacle tubes (Figure 8).<br />

Figure 8: Left lower grommet<br />

Instructions for Replacing the Housing:<br />

1. Place the device on its back side.<br />

2. Seat the feet of the front housing into the corresponding grooves on the chassis (Figure 9)<br />

then push down and toward the device to secure the housing in the chassis.<br />

Figure 9: Housing feet and corresponding chassis grooves<br />

3. Replace the two screws that connect the front housing to the chassis (Figure 2). Tighten<br />

screws to 6 in-lbs.<br />

4. Return the concentrator to the upright position<br />

96-03997-00-01 revA Service <strong>Manual</strong>, <strong>Inogen</strong> <strong>One</strong> <strong>G3</strong><br />