Figure 10: Screws to be removed from motherboard<br />
3. Carefully pull the motherboard about a half inch away from the device, making sure not<br />
to bend it in any way. With the cables and tubing exposed, gently unplug the two feed<br />
waste manifold cables, the product manifold cable, and the fan cable from the<br />
motherboard (Figure 11).<br />
Product manifold cable<br />
Fan cable<br />
Feed waste cables<br />
Figure 11: Disconnect all cables that connect device to motherboard<br />
4. Unplug the oxygen sensor tubing from the product manifold by pulling the tube off of the<br />
barb. Also unplug the accumulator tubing from the accumulator cap by pushing down on<br />
the retaining ring of the quick connect while pulling gently on the tubing (Figure 12).<br />
Pulling the tube too hard will result in breaking the pressure sensor free from the circuit<br />
board.<br />
96-03997-00-01 revA Service <strong>Manual</strong>, <strong>Inogen</strong> <strong>One</strong> <strong>G3</strong><br />