Chapter 5: <strong>EQAS</strong> ReportsTemplate 1This template includes all possible categories reported on. It includes a header with the <strong>Program</strong> Name andCatalog Number, Cycle Number, Sample Number and Sample Date. The data columns start from the left andinclude: Lab Number, Lab Name, Analyte name, Instrument name, Instrument Code, Method name, MethodCode, Reagent name and Reagent Code, Result Value and the unit submitted, a Converted Result andassociated unit (for an easy comparison of values within the subgroup), Comparator level, ComparatorZ-score, Comparator RMZ, Peer N, Peer Mean, Peer SD, Peer CV, Peer U, Method N, Method Mean, MethodSD, Method CV, Method U, Mode N, Mode Mean, Mode SD, Mode CV, Mode U and any UnreportableResults (UR) Codes submitted. Results outside acceptable limits (2.00 |Z-score|) or missing results arehighlighted in bold red text.72 <strong>EQAS</strong> <strong>Program</strong> <strong>User</strong> <strong>Guide</strong>