5.4.2 Inspecting the water installation<br />

32<br />

Use the following check list to ascertain that the installation has been performed correctly:<br />

– Water supply<br />

Has filter valve or shut-off valve and filter 5 µm respectively been installed in supply line?<br />

Have admissible water pressure (1.0 – 10 bar) and temperature (1 – 40 °C) been observed?<br />

Does supply capacity match the humidifier(s)?<br />

Are all pipes properly secured (threaded connections tightened)?<br />

Is the feed pipe properly sealed?<br />

– Water drain<br />

Has minimum inside diameter of drain pipes been maintained at least 40 mm throughout the<br />

entire length?<br />

Has drain pipe been installed with a downslope of at least 10 %?<br />

Has the heat resistance of the material used been verified to be at least 100°C?<br />

Are hoses and lines properly secured (hose clamps and threaded connections tightened)?