6.6.2 Dismantling and re-assembly work<br />

54<br />

Note: The lime collector tank must be replaced every 5000 hours, or after 3 years at the latest.<br />

Removal of the lime collector tank<br />

Caution! Before commencing any dismantling work, the steam-air humidifier is to be taken out<br />

of operation as described in chapter 6.3 and rendered safe from inadvertent operation.<br />

1 2<br />

1. Empty the lime collector tank via the drain valve.<br />

Caution: Danger of scalding! The water in the lime collector tank may be hot.<br />

2. Press the securing knob on both rotation locks. Then rotate both rotation locks clockwise and remove<br />

the lime collector tank downwards, from the base of the unit.<br />

Installation of the lime collector tank<br />

Installation of the lime collector tank is carried out in the reverse sequence to that detailed above for<br />

removal.<br />

Important! Check O-rings in the tank coupling sleeve of the unit and replace if defective. Before<br />

remounting the lime collector tank, grease the O-rings in the tank coupling sleeve with silicone grease.<br />

Important! Turn the rotation locks anticlockwise up to the stop (both securing knobs must engage).<br />

Important! Close drain valve on lime collector tank.