No. of Poles<br />

(See Circuit<br />

Diagram)<br />

Cross Section of Core Total Diameter<br />

5 0.25…0.38 mm 2 5.2…5.7 mm<br />

10 0.25…0.38 mm 2 7.7…8.2 mm<br />

(v) Maintenance - In the interest of measuring accuracy, it is advisable to check out<br />

the display and to regenerate the hairs in saturated moist air from time to time.<br />

However, if there are only minor changes in humidity, it is sufficient to re-set the<br />

hairs, regeneration isn't necessary. Hygro-transmitters used in moist air with<br />

over 80% relative humidity do not require regeneration. Transmitters which are<br />

exposed to dry air frequently or over long periods of time must be regenerated<br />

more frequently. Hygro-transmitters which have not been regenerated will, in the<br />

course of time, start to indicate values which are about 5-8% too high. This<br />

degeneration can be reserved by exposing the hygro-transmitter to saturated<br />

moist air.<br />

The easiest way to do this is to wrap the immersion stem of the hygro-transmitter<br />

for about 60 minutes in a damp cloth which has been dipped in lukewarm water.<br />

Make sure that the stem is completely enclosed in the cloth and that no air can<br />

get at it.<br />

Normally, the indicator will settle at 94-96% relative humidity. If, there are<br />

significant deviations from this value, then correct this by means of the setting<br />

screw marked in red at the lower end of the stem. The hygro-transmitter which is<br />

correct at 95% relative humidity will also be correct and in accordance with the<br />

adjustment at all other humidities, provided that the characteristic hygrometric<br />

features of the hair have not changed as a result of damaging influences (heat,<br />

aggressive vapors, mechanical tearing etc.).<br />

If, the hygro-transmitter is used with a wind protector, then it will be necessary to<br />

check this regularly. If, it is dirty, remove it from the instrument and clean it.<br />

3.1.2 Temperature<br />

(i) Hygro-thermo-transmitter - Hygro-thermo-transmitters resemble the hygrotransmitters<br />

just described. They are equipped with an additional hard-glass<br />

resistor Pt 100 in the immersion stem. This resistor is suitable for use when<br />

long-range measurements of temperature with high resolutions are required.<br />

The change in resistance in the platinum coiling as a result of changes in<br />

temperature is a measure of temperature. The relationship between resistance<br />

59