9.1.3.2 Solution B - Pipet 25 ml of solution A into a 250 volumetric flask and<br />

dilute to volume with absorbing solution. This contains 1 µg NO - 2/ml. Prepare<br />

fresh daily.<br />

9.2 Calibration<br />

9.2.1 Flowmeter - Calibrate the flowmeter against a calibrated flow measurement<br />

standard, such as a wet test meter, bubble flowmeter, or other reliable volume<br />

measurement standard. Calibrate in units of standard cm 3 /min (i.e., corrected to<br />

25 o C and 760 torr).<br />

9.2.2 Absorber - Calibrate the polypropylene absorber (see 8.1.2) by pipeting 50 ml of<br />

water or absorbing reagent into the absorber. Scribe the level of the meniscus<br />

with a sharp object, mark over the area with a felt-tip marking pen, and rub off the<br />

excess.<br />

9.2.3 Spectrophotometer<br />

9.2.3.1 Prepare calibration curve using 1 µg/ml working standards (see 9.1.3.2).<br />

9.2.3.2 In accordance with the analytical procedure given in 9.4, measure and<br />

record the absorbance for each calibration standard (0, 1, 2, 3, 4, 5, 6, 7,<br />

8, 9, 10, 12, 15, 20 µg NO2.<br />

9.2.3.3 Plot absorbance (y-axis) versus the corresponding concentration in µg<br />

NO - 2/50 ml final solution (x-axis). Draw or compute the straight line best<br />

fitting the data to obtain the calibration curve.<br />

9.3 Sample Collection<br />

9.3.1 4-Hourly Sampling<br />

9.3.1.1 Assemble the sampling apparatus (Figure 1) at the sampling site.<br />

Components upstream from the absorber may be connected, where<br />

required, with Teflon tubing; glass tubing with dry ball joints; or glass<br />

tubing with butt-to-butt joints with Teflon or polypropylene.<br />

9.3.1.2 Add exactly 30 ml of absorbing reagent to the calibrated absorber.<br />

9.3.1.3 Disconnect the funnel, connect the calibrated flowmeter, measure the flow<br />

rate before sampling and record as F1. If, the flow rate before sampling is<br />

not 1 Lpm, replace the flow control device and/or check the system for<br />

leaks. Start sampling only after obtaining an initial flow rate of 1 Lpm.<br />

9.3.1.4 Sample for 4 hrs. Record the exact sampling time in minutes at ts.<br />

124