9.0 REAGENTS<br />

9.1 Purity - All reagents shall conform to specifications for reagent grade chemicals<br />

of the American Chemical Society.<br />

9.2 Zero <strong>Air</strong> - <strong>Ambient</strong> air purified by passing through an activated charcoal filter, or<br />

by another appropriate manner (13).<br />

10.0 PROCEDURE<br />

10.1 Operate any commercial ozone analyzer following procedures given in the<br />

manufacturer's manual. Perform the calibration as directed in Section 11.0.<br />

Upon completion of a satisfactory calibration the analyzer is acceptable for ozone<br />

monitoring.<br />

11.0 CALIBRATION<br />

11.1 Connect the ozone analyzer to the output manifold of the calibration system<br />

(Figure-3). Check to insure proper operating parameters according to the<br />

instrument manual.<br />

11.2 Connect either a UV photometer (Section 8.4.2) to the output manifold. Check to<br />

insure proper operating parameters according to the instrument manuals. Either<br />

of these instruments can determine the true ozone concentration of the<br />

calibration air streams.<br />

11.2.1 As an alternative, the ozone concentration of the calibration air streams may be<br />

determined by using the Neutral buffered Potassium Iodide method.<br />

11.3 Zero <strong>Air</strong> - With the O3 lamp off, flush the system for about ten minutes to remove<br />

residual O3. While the analyzers sample the zero air, record ten consecutive<br />

digital display values. Calculate, for the ozone analyzer being calibrated and<br />

either the UV photometer or ozone transfer standard, the sum and average of the<br />

ten values and record. If an ozone transfer standard is used, note and record<br />

temperature and pressure.<br />

11.3.1 If the spectrophotometeric procedure is used, a manual sample of the zero air<br />

should be taken. Record the ppm of ozone found, performing the calculations<br />

described in the procedure.<br />

11.4 Test Atmospheres - With the O3 lamp well equilibrated, adjust the O3<br />

concentration to the range desired for a calibration point. Wait 10 minutes for<br />

equilibration at each setting. Measure and record the analyzer digital outputs as<br />

in the preceeding Section 11.3. Test atmospheres of approximately 80, 60, 40<br />

and 20% of the range of the analyzer should be run for a calibration.<br />

49