prior to analysis. The interference of trace metals may be eliminated by the<br />

addition of ethylenediamine tetra acetic acid (EDTA) to the absorbing solution<br />

prior to sampling. At least 60 µg iron (III), 10 µg manganese (II), and 10 µg<br />

chromium (III) in 10 ml absorbing reagent can be tolerated in the procedure. No<br />

significant interference was found from 10 µg copper (II) and 22 µg vanadium<br />

(V). Ammonium, sulphide, and aldehydes do not interfere.<br />

6.0 APPARATUS<br />

6.1 Sampling<br />

6.1.1 Absorber - An all-glass midget impinger, as shown in Figure-1, is recommended<br />

for 30 minutes, 1 hour and 4 hours samples. For 24 hours sampling, assemble<br />

an absorber from the following parts.<br />

6.1.3 Polypropylene Two-Port Tube Closures<br />

6.1.3 Glass Impingers - Tubing, 6 mm OD and 15 cm long. One end is drawn to<br />

small diameter so that a No.79 jeweller's drill bit will pass through, but a No.78<br />

jeweller's bit will not. The other end is fire polished.<br />

6.1.4 Polypropylene Tubes - Tubes 164 by 32 mm, 'Nalgene' or equivalent.<br />

6.1.5 Pump - Capable of maintaining an air pressure differential greater than 0.7 atm<br />

at the desired flow rate.<br />

6.1.6 <strong>Air</strong> Flowmeter or Critical Orifice - A calibrated rotameter or critical orifice<br />

capable of measuring air flow within 2%. For 30 minutes sampling, a 22 gauge<br />

hypodermic needle 2.5 cm long may be used as a critical orifice to give a flow of<br />

about 1 litre/minute. For 1 hour sampling, a 23 gauge hypodermic needle 1.6 cm<br />

long may be used to give a flow of about 0.2 litre/minute. Use a membrane filter<br />

to protect the orifice (Figure 2).<br />

6.2 Analysis<br />

A spectrophotometer suitable for measurement of absorbance at 560 nm with an<br />

effective spectral band width of less than 15 nm is required. Reagent blank<br />

problems may occur with spectrophotometer having greater spectral band<br />

widths.<br />