Operating voltage<br />

Electronic : + 15 V/60 mA<br />

Heater : 24 V/3 A<br />

Connection : 19 pol. Euchner plug<br />

Case material : GD-A1, sea waterproof<br />

Transmission line<br />

lon<br />

: LiCY 16x0.75, max. 0 20 mm, 100 m<br />

Recomm. mast dimensions : Threaded pipe 1 1/2" in accordance with<br />

DIN 2441<br />

Dimensions : See scale drawing (Figure-3).<br />

Wind stress at 35 m/s : Approx. 5 kp<br />

Weight : 3.2 kg<br />

(iv) Set-up and Mode of Operation - Wind direction is detected by means of a<br />

highly sensitive light metal wind vane whose ball-bearing pivot is connected to a<br />

code drum. The casing of the code drum has a code pattern which can be<br />

scanned opto-electronically. The resolution of the code pattern amounts of 144<br />

steps per 360 o . An 8-bit binary information has been used i.e. the code drum has<br />

8 scanning tracks. The Gray Code which counts up from 0...71 and down from<br />

71...0 has been utilized. Seven tracks are required for this while the 8th track<br />

serves to differentiate between up and down. The scanning of the code pattern<br />

is carried out by means of optical reflective sensors where a diode emits an<br />

infrared ray which is reflected by the drum casing surface onto a photo-transistor.<br />

The output voltage of the transistor depends on the position of the drum (the<br />

degree of reflection). The degree of reflection of the drum background is less<br />

than of the code pattern bus is still nonzero. A drop-page of the signal voltage<br />

below the lower trigger level indicates a malfunction of the corresponding<br />

scanner component and a warning signal is given. Alternatively a potentiometer<br />

direction sensor can be used where the wind vane is coupled to a 5 K<br />

potentiometer.<br />

Opto-electronic semiconductor components are influenced by temperature<br />

fluctuations and by ageing. Thus, the diode current in the sensor must be<br />

adjusted so as to compensate for interference. For this purpose, an additional<br />

continuous tract is scanned and the output voltage is kept constant with the<br />

regulated diode current. If, transmission factor changes due to the influences of<br />

temperature or ageing, this will be automatically taken into account and the diode<br />

current will be readjusted. Wind velocity is recorded by means of a highly<br />

sensitive rotating cup anemometer whose ball bearing pivot is connected to a<br />

slotted drum. The slotted drum is also scanned opto-electronically.<br />

The impulse frequency emitted is proportional to wind velocity. Just as was the<br />

case for the wind direction part, the output signal of the velocity sensor is<br />

electronically monitored to ensure that there is no malfunction of the scanner.<br />

(Installation of lightning rod No. 3100.99 is recommended in areas with frequent<br />

lightning activity).<br />

64