section comprises two independent flow controls that can be varied individually to<br />

provide a dilution ration of up to 1,000 to 1. The GPT section comprises a<br />

current-regulated ozone generator through which a portion of the dilution air<br />

flows even when the ozonator is not in operation. For dynamic dilution, the<br />

metered NO combines with this portion of the dilution air and passes through the<br />

reaction chamber. It then combines with the balance of the dilution air and<br />

passes through the sampling manifold. For GPT the flow path is the same<br />

except that a portion of the oxygen in the air passing through the ozone<br />

generator is converted to ozone.<br />

10.2.1 <strong>Air</strong> Flow Controller - A device capable of maintaining constant clean-air flow up<br />

to 5 L/min within + 2% of the required flow rate.<br />

10.2.2 <strong>Air</strong> Flowmeter - A calibrated flowmeter capable of measuring air flow rates<br />

within + 2%.<br />

10.2.3 Nitric Oxide Flow Controller - A device capable of maintaining constant NO<br />

flow within + 2% of the required flow rate.<br />

10.2.4 Nitric Oxide Flowmeter - A calibrated flowmeter capable of measuring NO flow<br />

rates within + 2%.<br />

10.2.5 Two-Stage Regulator - The two-stage pressure regulator for the standard NO<br />

cylinder must be of stainless steel to prevent any reaction of the NO gas.<br />

10.2.6 Ozone Generator - The generator must be capable of generating stable levels of<br />

O3 for the GPT of NO to provide NO2 concentrations throughout the calibration<br />

range.<br />

10.2.7 Reaction Chamber - The chamber used for the reaction of O3 with excess NO<br />

should be of sufficient volume that the residence time is not less than 2 minutes<br />

(11).<br />

10.2.8 Mixing Chamber - A chamber used to allow thorough mixing of reaction<br />

products and dilution air.<br />

10.2.9 Sampling Manifold - The sampling manifold should be of adequate design to<br />

insure against any pressure buildup. It must have a vent to insure atmospheric<br />

pressure at the manifold, and for safety considerations to direct excess<br />

calibration air streams to an exhaust system.<br />

10.3 Procedure - Prior to start of calibration, for safety purposes, insure proper<br />

venting of the analyzer exhaust and the calibration system have excess flow.<br />

Insure that the analyzer and the calibration system have been on for a time<br />

sufficient to provide stable operation.<br />