(iii) Technical Data<br />

Measuring range : 10... 100% relative humidity<br />

Accuracy : + 2% relative humidity<br />

Length of scale : 94 mm<br />

Graduation : 1% relative humidity<br />

Load : Maximum 2 W<br />

Slider current : Maximum 100 mA<br />

<strong>Ambient</strong> temperature : -35... + 70 o C<br />

Weight : 0.45 kg.<br />

Models Available<br />

Electrical outlet : Simple<br />

Double<br />

Measuring element : H<br />

S (Hygrofix)<br />

Type of connection : Lemosa plug<br />

3 m cable<br />

Electrical output : 40-50-40 ohms<br />

50-30-50 ohms<br />

0...100 ohms<br />

50-150-50 ohms<br />

0...200 ohms<br />

0...1000 ohms<br />

0...2000 ohms<br />

0...200 ohms lin.<br />

(iv) Installing the Instrument - Install the hygro-transmitter in such a way that it is<br />

protected from jarring, dust, chemical impurities and splashing water. It can be<br />

mounted with the aid of the R 3/4" nuts directly to the wall of the room where the<br />

measurements are to be taken or can use the enclosed mounting angle to screw<br />

it to a plane surface. Please make sure that the transmitter is in the correct<br />

position for use - the immersion stem must be vertical and pointing downwards.<br />

Should there be heavily polluted air or wind velocities above about 3 m/s, then it<br />

will be necessary to employ a wind protection device in order to guarantee<br />

accurate measurements. This device is simply slipped over the immersion stem<br />

of the hygro-transmitter and turned so that the closed side faces the air stream.<br />

Connect the electrical output of the hygro-transmitter to an appropriate display<br />

instrument (see circuit diagram). If, the hygro-transmitter has a double outlet, the<br />

second outlet can be used to connect a recorder or an additional display<br />

instrument. If, the hygro-transmitter has a Lemosa plug connection, then we<br />

recommend using a flexible pilot wire LiYY in the following models for the<br />

electrical connection :<br />