11.4.1 If the spectrophotometeric procedure is used, a manual sample is taken of each<br />

calibration point. Record the ppm of ozone found for each calibration level by<br />

performing the calculations described in the procedure.<br />

12.0 CALCULATIONS<br />

12.1 If a UV photometer was used in the calibration, the ozone readings are the true<br />

ozone concentrations already corrected to standard conditions (25 o C and 101.3<br />

kPa).<br />

Where :<br />

True Ozone (ppm) = Ozone Reading - Zero Reading<br />

Ozone Reading = The UV photometer ozone readout for each calibration point<br />

test atmosphere (11.4)<br />

Zero Reading = The UV photometer ozone readout for the zero air stream<br />

(11.3)<br />

12.2 If a transfer standard was used in the calibration, its ozone readings must be<br />

corrected to standard conditions (25 o C and 101.3 kPa).<br />

Ozone Zero P 298<br />

True Ozone (ppm) = Reading - Reading X ------- X ------------<br />

101.3 t + 273<br />

Where :<br />

Ozone Reading = the transfer standard readout for each calibration, point test<br />

atmosphere (11.4)<br />

Zero Reading = the transfer standard ozone readout for the zero air stream<br />

(11.3)<br />

P = the barometric pressure during the calibration in kPa (11.2)<br />

t = the temperature during the calibration in o C (11.2)<br />

12.3 If the spectrophotometric procedure was used in the calibration the true ozone is<br />

calculated as follows :<br />

True Ozone (ppm) = Ozone Conc.CTA - Ozone Conc.ZAS<br />