15.0 OPERATIONAL CHECKS<br />

15.1 Zero and Span Settings - If the required zero and span corrections performed in<br />

accordance with clause 14.2 are greater than 80% of the range, have the<br />

analyser serviced.<br />

15.2 Sample Flow Rate - If the sample flow rate has changed by more than + 20% of<br />

the initial value, check the particulate filter for blockage, and the sample pump for<br />

proper operation. Check the filter monthly by measuring the flow rate with and<br />

without the filter in place. Replace the filter if the drop is more than 5%.<br />

15.3 Temperature Control - Check the temperature of the shelter or room in which<br />

the analyser is located. If, it has changed by more than + 5 o C, have the heatingcooling<br />

system serviced.<br />

15.4 Analog Recording System - Check the strip chart recorder for the following :<br />

(a) Chart speed setting<br />

(b) Gain control setting<br />

(c) Ink supply<br />

(d) Paper supply<br />

(e) Excessive noise<br />

(f) Proper operation of analyser<br />

(g) Time-mark, sign and date the record<br />

15.5 Digital Recorder - Perform the checks in accordance with the manufacturer's<br />

instructions.<br />

16.0 REFERENCES<br />

1. Dechant, R.F. and Mueller, P.K., "Performance of Continuous NDIR Carbon<br />

Monoxide Analyser", Report No. 57, <strong>Air</strong> and Industrial Hygiene Laboratory,<br />

Department of Public Health, Berkeley, CA, USA, June 1969.<br />

2. Jacobs, M.B., Braverman, M.M. and Hochheiser, S., "Continuous Determination<br />

of Carbon Monoxide and Hydrocarbons in <strong>Air</strong> by a Modified Infrared<br />

Spectrometer", J. <strong>Air</strong> <strong>Pollution</strong> Control Assoc., 9, 1959, pp. 110-114.<br />

3. Moore, M.A., "Critical Evaluation of the Analysis of Carbon Monoxide with<br />

Nondispersive Infrared, NDIR", presented at the Ninth Conference on Methods in<br />

<strong>Air</strong> <strong>Pollution</strong> and Industrial Hygiene Studies, Passadena, CA, USA, Feb. 7-9,<br />

1968.<br />

4. McKee, H.C. and Childers, R.E., "Collaborative Study of Reference Method for<br />

the Continuous Measurement of Carbon Monoxide in the Atmosphere<br />