11. CONTINUOUS MEASUREMENT OF OZONE IN THE<br />

ATMOSPHERE BY ULTRAVIOLET PHOTOMETRIC<br />

INSTRUMENTS<br />

1.0 TITLE<br />

Method for continuous measurement of Ozone in the atmosphere by Ultraviolet<br />

Phtometric Instruments.<br />

2.0 PURPOSE<br />

The purpose is to lay down a uniform and reliable method for measurement of<br />

ozone in ambient air.<br />

3.0 PRINCIPLE<br />

3.2 The method is based on the photometric assay of ozone (O3) concentrations in a<br />

dynamic flow system. The concentration of O3 is determined in an absorption<br />

cell from the measurement of the amount of light absorbed at a wavelength of<br />

254 nm. The method is based on the absorption coefficient of O3 at 254 nm, the<br />

optical path length through the sample, and the transmittance, temperature and<br />

pressure of the sample (1,2,3,4,5). The quantities above are related by the Beer-<br />

Lambert absorption law.<br />

I<br />

Transmittance = --- = e -αc1<br />

Io<br />

Where :<br />

α = absorption coefficient of O3 at 254 nm = 310 atm -1 cm -1 at 0 o C and<br />

101.3 kPa<br />

e = O3 concentration in units of atmosphere<br />

1 = optical path of absorption cell length in cm<br />

I = intensity of light passing through cell with an ozone sample<br />

Io = intensity of light passing through cell with zero air<br />

Typically, an air sample is first directed through a scrubber that removes any O3<br />

present, but otherwise does not affect the sample. The ozone-free sample then<br />

flows through the absorption cell, and its transmittance is measured. This<br />

constitutes the zero cycle. At a present time, solenoid switches and another air<br />

sample flows directly into the absorption cell, bypassing the scrubber and its<br />

transmittance is measured. This constitutes the ozone measurement cycle. The<br />

difference in transmittance between the two cycles is a measure of the O3<br />

concentration. The complete measurement cycle takes about 20 to 30 s.<br />

146