Checks and Preparation<br />

2.12 Noise Countermeasures<br />

Be sure to connect the chassis GND of the object to be measured.<br />

It may become effective by ensuring that the chassis GND wire of the measurement object is<br />

connected to a good ground.<br />

Measurement object<br />

Thermocouple<br />

R1<br />

Input terminals<br />

+<br />

<strong>GL900</strong><br />

Vin<br />

R2<br />

–<br />

Z3<br />

Z1<br />

Z2<br />

Connect the signal chassis GND and the measurement device chassis ground.<br />

Use a short, thick lead to connect the chassis GND of the measurement object to the <strong>GL900</strong>’<br />

chassis GND. It will become even more effective if the ground potentials are the same.<br />

Measurement device chassis<br />

<strong>GL900</strong><br />

GND<br />

GND<br />

Examples of noise countermeasures<br />

If measured values fluctuate due to external noise, the following countermeasures are<br />

recommended.<br />

(Results may differ depending on the noise type.)<br />

Example 1: Connect the <strong>GL900</strong>'s GND to ground.<br />

Example 2: Connect the <strong>GL900</strong>'s GND to the measurement target's GND.<br />

Example 3: In the AMP Settings menu, set Filter to any setting other than OFF.<br />