26.12.2014 Views

Fisher® FIELDVUE DVC6200 Digital Valve Controller

Fisher® FIELDVUE DVC6200 Digital Valve Controller

Fisher® FIELDVUE DVC6200 Digital Valve Controller

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Instruction Manual<br />

D103605X012<br />

Configuration<br />

July 2012<br />

Control Mode<br />

Control Mode—This displays the current control mode of the instrument. This will show Analog if the instrument is<br />

in Point­to­Point mode and is using a 4­20 mA signal for its power and set point. This will show <strong>Digital</strong> if the<br />

instrument is in Multidrop mode and is using 24 VDC for power and a digital set point for control.<br />

Note<br />

Another mode, Test, may be displayed. Normally the instrument should not be in the Test mode. The digital valve controller<br />

automatically switches to this mode whenever it needs to stroke the valve during calibration or stroke valve, for example.<br />

However, if you abort from a procedure where the instrument is in the test mode, it may remain in this mode. To take the<br />

instrument out of the Test mode, select Change Control Mode and enter Analog or <strong>Digital</strong>.<br />

Change Control Mode—This allows the user to configure the control mode to Analog or <strong>Digital</strong>.<br />

Restart Control Mode—This defines the Control Mode of the instrument after a restart (e.g. power cycle). Available<br />

choices are Resume Last, Analog and <strong>Digital</strong>.<br />

Characterization<br />

Input Characterization<br />

Input Characterization defines the relationship between the travel target and ranged set point. Ranged set point is the<br />

input to the characterization function. If the zero power condition equals closed, then a set point of 0% corresponds to<br />

a ranged input of 0%. If the zero power condition equals open, a set point of 0% corresponds to a ranged input of 100%.<br />

Travel target is the output from the characterization function.<br />

To select an input characterization, select Input Characterization from the Characterization menu. You can select from<br />

the three fixed input characteristics shown in figure 3‐2 or you can select a custom characteristic. Figure 3‐2 shows the<br />

relationship between the travel target and ranged set point for the fixed input characteristics, assuming the Zero<br />

Power Condition is configured as closed.<br />

You can specify 21 points on a custom characteristic curve. Each point defines a travel target, in % of ranged travel, for<br />

a corresponding set point, in % of ranged set point. Set point values range from -6.25% to 106.25%. Before<br />

modification, the custom characteristic is linear.<br />

Custom Characterization<br />

To define a custom input character, from the Characterization menu select Custom Characterization. Select the point<br />

you wish to define (1 to 21), then enter the desired set point value. Press Enter then enter the desired travel target for<br />

the corresponding set point. When finished, select point 0 to return to the Characterization menu.<br />

With input characterization you can modify the overall characteristic of the valve and instrument combination.<br />

Selecting an equal percentage, quick opening, or custom (other than the default of linear) input characteristic<br />

modifies the overall valve and instrument characteristic. However, if you select the linear input characteristic, the<br />

overall valve and instrument characteristic is the characteristic of the valve, which is determined by the valve trim (i.e.,<br />

the plug or cage).<br />

21

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!