NanoLC 1D Plus and 2D System Operator's Manual - Eksigent

NanoLC 1D Plus and 2D System Operator's Manual - Eksigent

NanoLC 1D Plus and 2D System Operator's Manual - Eksigent

SHOW MORE
SHOW LESS
  • No tags were found...

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

5.4 Creating an Autosampler MethodQuick-start Guide..37The parameters used for loading the sample into the injection valve <strong>and</strong> for rinsing the autosamplersyringe <strong>and</strong> sample needle are stored in the autosampler method. This section will review anautosampler method appropriate for loading a trap with Channel 1, <strong>and</strong> running a gradient with Channel2.step 1 Place the sample vial containing the st<strong>and</strong>ard test mixture in vial position A1 of autosampler 48-vialtray.step 2 Click Run Manager in the Acquisition Window to open the Run Manager display (Figure 5-6).Figure 5-6. Run Manager Windowstep 3 If you do not see a picture of the 48-vial tray in the Run Manager Window select the Devices >Autosampler Device Settings menu to set the tray type. You will need to restart the Run Managerafter changing the tray type.step 4 If you do not see the columns shown in Figure 5-6, choose Edit > Choose Column… from themenu <strong>and</strong> select the appropriate columns for display.step 5 Click Autosampler Methods to bring up the Autosampler Method Editor Window (Figure 5-7).nanoLC Operator’s <strong>Manual</strong> 611-00012 Revision A1