52 Operator’s <strong>Manual</strong>6.4 Troubleshooting Checklist<strong>System</strong> Initialization…Table 6-1. Troubleshooting ChecklistSymptom Possible Cause Corrective ActionPower LED on front panel is noton.Power cord not connectedNo power at outletPower LED has failed but systemresponse OKConnect power cord.Repair electrical outlet.Contact <strong>Eksigent</strong> TechnologiesTechnical Support for assistance.Front panel power LED is on butsoftware fails to recognize theinstrument’s presence.Communication error betweencomputer <strong>and</strong> an LC systemVerify that the instrument’s serial cableis securely connected to the computer’sCOM1 serial port. (see Section2.5) Reboot computer <strong>and</strong> cycle poweron instrument.Contact <strong>Eksigent</strong> TechnologiesTechnical Support for assistance.Liquid leaking at the bottom of themobile phase reservoirsThe reservoirs were not placed inthe pump adapter properlyVerify the stopcock is placed properly<strong>and</strong> tightly screwed in.Loud hissing sound from theinstrumentAir leaks from the air inlet fittingVerify the air tubing is connectedproperly to the gas fitting.Tighten the air inlet gas fitting.Contact <strong>Eksigent</strong> TechnologiesTechnical Support for assistance.nanoLC Operator’s <strong>Manual</strong> 611-00012 Revision A1