12.07.2015 Views

NanoLC 1D Plus and 2D System Operator's Manual - Eksigent

NanoLC 1D Plus and 2D System Operator's Manual - Eksigent

NanoLC 1D Plus and 2D System Operator's Manual - Eksigent

SHOW MORE
SHOW LESS
  • No tags were found...

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

50 Operator’s <strong>Manual</strong>6.2 Calibration ValuesThe Calibration Values tab of the Hardware Diagnostics Window summarizes the current K values ofthe flow meter, the gain <strong>and</strong> zero offsets of pressure transducers <strong>and</strong> the PID parameters for thepumps. The following settings are typical for a nanoLC <strong>and</strong> are shown for example only. The window onthe left is for channel 1 <strong>and</strong> the window on the right is for channel 2,Figure 6-1. Hardware Diagnostics — Calibration Values TabThe values should be documented before <strong>and</strong> after as part of maintaining a good instrument log, <strong>and</strong>noting changes. These values should be very similar from diagnostic to diagnostic; large changes canindicate a problem.nanoLC Operator’s <strong>Manual</strong> 611-00012 Revision A1

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!