13.07.2015 Views

Artis zee/zeego VC14 - Siemens Healthcare

Artis zee/zeego VC14 - Siemens Healthcare

Artis zee/zeego VC14 - Siemens Healthcare

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

s<strong>Artis</strong> <strong>zee</strong>/<strong>zee</strong>go <strong>VC14</strong>DICOM Conformance Statement4.2.6.3.2.3 SOP Specific ConformanceThe Printer SOP Class allows to monitor the status of the hardcopy printer in a synchronous andan asynchronous way.The Print SCU AE application will cyclically “ask” the Printer (SCP) for its status synchronously:• N-GET as SCUThe following information is supported:Table 38 - Used Printer N-EVENT Report attributesEvent-type Name Event Attributes Tag Usage SCUNormal 1Warning 2 Printer Status Info (2110,0020) UFailure 3 Printer Status Info (2110,0020) UU = User OptionTable 39 - Mandatory Printer N-GET-RSP, N-EVENT-REPORT-RQ attributesAttribute Name Tag Usage SCP Supported ValuesPrinter Status (2110,0010) M NORMAL, FAILURE, WARNINGPrinter Status Info (2110,0020) M See tables in Annex for details.M = MandatoryNote: For a detailed description on how <strong>Artis</strong> <strong>zee</strong>/<strong>zee</strong>go reacts on different printer status messages, please refer to theAnnex section “DICOM Print SCU - detailed status displays”.4.2.6.4 Association Acceptance Policyn. a.© <strong>Siemens</strong> AG, <strong>Healthcare</strong> Sector 2009 Rev. 08 Page 44 of 105