SECTION S | SOIL<br />

PORE PRESSURE TRANSDUCER<br />

It is a good reliability electronic device used to measure pore<br />

pressure. It requires a de-airing block. Every transducer must be<br />

connected to the automatic data acquisition system mod. S334<br />

(see p. 559) or mod. S301-05 ( see p. 544).<br />

- Input voltage: 10 volts dc, Sensitivity range: 2 - 4 mV/V<br />

- Accuracy: 0.15 fs<br />

- Pressure connection: 0.25 BSP<br />

- Protected against corrosive pore water pressure<br />

- 2 metres cable and 5 pin plug included<br />

MODELS<br />

S336-50 Pressure transducer up to 1000 kPa<br />

S336-51 Pressure transducer up to 2000 kPa<br />

ACCESSORIES<br />

S336-55 De-airing block for pressure transducer<br />

S336-50<br />

EXTENSION CABLE FOR TRANSDUCERS<br />

MODELS<br />

S336-30 2 metres long<br />

S336-31 5 metres long<br />

S336-32 10 metres long<br />

S336-55<br />

S336-51<br />

S356N<br />

2 CHANNELS UNIT FOR PORE<br />

PRESSURE MEASUREMENT<br />

Based on a high resolution graphic display this instrument allows<br />

continuous control and monitoring of pore pressure measurements<br />

coming from pressure transducers.<br />

Models like S336-50 or S336-51 supplied along with proper connection<br />

cable can be plugged into this new S356N readout so that<br />

manual triaxial tests can be performed without need of automatic<br />

data acquisition units or outdated mercury manometers.<br />

S356N unit allows real time visualization of the pore pressure<br />

measurements and subsequent peak data recording through its<br />

dedicated peak function.<br />

Technical specifications<br />

- Measuring units (selectable): mbar, bar, MPa, kPa, psi<br />

- Acquisition and data processing system at 24 bit, effective resolution:<br />

17 bit.<br />

- Operator interface composed by 5 multi-functions pushbuttons;<br />

function icons shown on the display<br />

- The two analogue-digital channels accept sensors at 2mV/V<br />

- Languages: see C108N at page 219...221<br />

- Graphic display with high resolution: 192x64 pixels<br />

- Class: 0.5% starting from 10% of maximum value<br />

Power supply: 230V 1ph 50-60Hz<br />

Dimensions: 230x145x240 mm<br />

Weight: 4 kg approx<br />

ACCESSORIES for S356N<br />

NEW<br />

- Pressure transducers, available models see p. 549<br />

- De-airing block<br />

- Extension cables for transducers<br />

Note: It is recommended to use not more than 10m of<br />

extension cable to avoid noise problems that<br />

might occur.<br />

Channel configuration and calibration<br />

S356N+S336-50+S336-51