8. Press ―ENTER‖ function.<br />
9. In CAL.ENTER mo<strong>de</strong>, choose one of the following options: 2.1- > VALUE 1; 2.2 -> VALUE 2<br />
and 2.3 -> UNITS. Pressing ―ENTER‖ function the required option is selected.<br />
―DOWN‖ function (▼) : To move through the character list in a <strong>de</strong>scending or<strong>de</strong>r.<br />
―UP‖ function (▲) : To move through the character list in an ascending or<strong>de</strong>r.<br />
―ESC‖ function: To return to the previous character or previous menu in the display.<br />
―ENTER‖ function: To validate the edited character and move to the next character.<br />
10. Once the corresponding value to the pending load has been assigned, the value in the<br />
display will be of three digits plus a comma. Units must be edited such that the value<br />
represented and the units are coherent. Validate the value inserted pressing START.<br />
VALUE 1 will be memorised.<br />
11. Press ―ESC‖ to return to the previous menu in or<strong>de</strong>r to edit the units. Units must have no<br />
more than three characters.<br />
12. Press STOP and restart the crane again. Replace the load with another known load.<br />
13. Repeat the process again (steps 4 to 8) selecting VALUE 2 in the Calibration menu.<br />
14. Once finished the feedback calibration process, the system will be calibrated.<br />
At this point the system is ready to work normally. Until calibration is performed, data appearing in<br />
the display will be inconsistent.<br />
Calibration generates a linear interpolation of the possible values. This means that it can be used<br />
with all systems that generate a linear analogue outputs. For non linear systems the interpolation will<br />
not produce a correct display value. Calibration allows to obtain negative values.<br />
Example of calibration of a load cell:<br />
Enter into the calibration mo<strong>de</strong> and introduce the value corresponding to the hook without load (1 st<br />
value).<br />
Validate the value pressing START. Extinguish the transmitter.<br />
Start the system again and raise a known load.<br />
Extinguish the transmitter and enter again to the calibration mo<strong>de</strong> for the edition of the 2 nd value.<br />
Press START to validate the value.<br />
Press pushbutton 1 to edit the units.<br />
Press START to validate the units.<br />
Extinguish the transmitter.<br />
The equipment is already ready to work normally.<br />