448 CHAPTER 9<br />

Substituting the given values in the above equation, we obtain<br />

˙D(0.1 MeV) = 8.1 × 10 −10 Gy/s<br />

˙D(2.0 Mev) = 8.2 × 10 −10 Gy/s.<br />

The dose rates for the two radiation fields are about the same. A particle-fluxmeasuring<br />

instrument, however, such as a GM counter, would register about 20<br />

times more for the 0.1-MeV radiation than for the higher energy radiation. Since<br />

ionization chambers measure absorbed energy rather than individual particles and<br />

are reasonably energy independent (Fig. 9-17), we use ionization-chamber-type instruments<br />

when measuring radiation dose rates (unless a specific counter-type instrument<br />

is calibrated for a specific gamma-ray energy, in which case it may be used<br />

for dosimetry for that quantum energy only).<br />

As shown in Example 9.3, particle-counting survey instruments, such as GM<br />

counters and scintillation counters, are very energy dependent when used <strong>to</strong> measure<br />

radiation dose (Fig. 9-18). The GM counter’s response is much flatter (although<br />

still not flat) with the beta window closed than with the beta window open.<br />

The reason for this is that the thicker wall, represented by the closed beta window,<br />

absorbs more low-energy pho<strong>to</strong>ns than does the open window. This is the<br />

basis for energy-compensated GM detec<strong>to</strong>rs. An energy-compensated GM detec<strong>to</strong>r<br />

has a wall thick enough <strong>to</strong> flatten the energy response <strong>to</strong> within about ±15% of<br />

the response <strong>to</strong> 137 Cs, yet thin enough <strong>to</strong> detect pho<strong>to</strong>ns down <strong>to</strong> about 50 keV<br />

(Fig. 9-19).<br />

Figure 9-17. Energy response curve for a general-purpose ionization-chamber-type survey meter with a 300mg/cm<br />

2 -thick retractable beta shield. (Courtesy of Ludlum Measurements, Inc, Sweetwater, TX.)